Recall Z-2430-2015 ยท KXE Prosthesis, Wrist, Hemi-, Ulnar

Stryker Howmedica Osteonics Corp.

Sigmoid Notch, Radial Plate, Small, Sterile, Do Not Reuse, Rx Only. Intended for replacement of the distal radio-ulnar joint following ulnar head resection arthroplasty.

DateAug 20, 2015 (initiated Jun 24, 2015)
StatusTerminated
ReasonStryker is recalling rHead, uHead, Sigmoid Notch, Remotion, Radio Capitellum because packaging integrity (sterile barrier) of the packaging type KIT I may potentially be compromised by transportation.
Root causePackaging change control
ActionBranches/Agencies were notified by Stryker of this action by e-mails on June 24, 2015 and an Urgent Medical Device Removal Letter/Acknowledgement Response Form dated June 25, 2015 was sent to the attention of the Risk Manager.
Quantity16992 total
DistributionWorldwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K061146SBI ULNAR HEAD IMPLANT
Small Bone Innovations, Inc., Jul 19, 2006