ClearanceBrief
Recall Z-2429-2015 ยท JDB Prosthesis, Elbow, Semi-Constrained, Cemented

Stryker Howmedica Osteonics Corp.

Recon, Radial Implant, Standard Stem, Size 1, Sterile, Do Not Reuse, Rx Only. Intended for replacement of the proximal end of the radius.

DateAug 20, 2015 (initiated Jun 24, 2015)
StatusTerminated
ReasonStryker is recalling rHead, uHead, Sigmoid Notch, Remotion, Radio Capitellum because packaging integrity (sterile barrier) of the packaging type KIT I may potentially be compromised by transportation.
Root causePackaging change control
ActionBranches/Agencies were notified by Stryker of this action by e-mails on June 24, 2015 and an Urgent Medical Device Removal Letter/Acknowledgement Response Form dated June 25, 2015 was sent to the attention of the Risk Manager.
Quantity16992 total
DistributionWorldwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K102180RHEAD RADIAL HEAD EXTENDED STEMS
Small Bone Innovations, Inc., Dec 20, 2010