Recall Z-2262-2015 ยท HWT Template

Encore Medical, Lp

FOUNDATION KNEE SYSTEM NON STERILE Femoral A/P Cut Block - Type II REF 800-01-139

DateJul 27, 2015 (initiated Jun 24, 2015)
StatusTerminated
ReasonThe markings on the distal face of the instruments are mis-oriented by 180. This includes Anterior and Posterior, which the surgeon uses to correctly place the cut block on the resectioned face of the femur. The cut block is not symmetric, so correct orientation of the instrument in the A/P direction is critical to making proper anterior, posterior, and chamfer cuts to fit the femoral implant.
Root causeComponent design/selection
ActionDJO Surgical sent an " Urgent Field Safety Notice" dated June 30, 2015, to the affected customer. The recalling firm requested the customer return the affected product for a replacement. For further questions, please call (512) 834-6302.
Quantity10 units
DistributionUS Distribution to TX and GA.
FDA recordOpen on fda.gov