Recall Z-2256-2015 ยท JAK System, X-Ray, Tomography, Computed

Philips Medical Systems (Cleveland) Inc

Computed Tomography X-ray Systems Brilliance iCT SP. Intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.

DateJul 24, 2015 (initiated Feb 5, 2015)
StatusTerminated
Reason5 issues. Fast Get Ready incorrectly enabled during service procedures. Contrast annotation missing for some images when manual contrast was administered. Dot artifact present, intermittently, after startup. Heart Rate measured value incorrectly remains constant for some heart rate change conditions. DoseRight incorrectly enabled when the Exam Card specified disabling DoseRight. Tube heat predict
Root causeSoftware change control
ActionOn 2/5/2015 the firm sent Customer Information Letters to their customers.
Quantity2 units
DistributionWorldwide Distribution-US (nationwide) including the states of AR, AZ, CA, CT, DE, FL, GA, IA, IN, KS, LA, MA, MD, MI, MN, MO, MS, MT, NH, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, VT, WV & WY, and to the following countries: Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Cambodia, Canad