BIOTRONIK, Inc.

PSW 1203.U/1, PSW 1301.U, PSW 1307.U, PSW 1403.U, and PSW 1501.U Pacemaker Programmer software. Software approved for use with BIOTRONIK device programmers (trade names: Renamic and ICS 3000). Allows physicians to program devices to pace exclusively in the left ventricle.

DateAug 13, 2015 (initiated Jun 30, 2015)
StatusTerminated
ReasonVentricular packing: LV software programming versions for BIOTRONIK CRT-P and CRT-D devices are recalled because they are not approved for use in the US.
Root causeSoftware Manufacturing/Software Deployment
ActionBIOTRONIK sent the Important Product Information letter and the Acknowledgement of Important Product Information Form, dated June 20, 2015, to affected physicians with the following recommendations: -Review programming settings for your patients with CRT-P or CRT-D devices. -If you have a patient with a device programmed to Ventricular pacing: LV, arrange to have the patients device reprogrammed to an appropriate approved mode (i.e. "Ventricular pacing: RV" or "Ventricular pacing: BiV"). -Physicians are requested to sign the Acknowledgement of Important Product Information Form and immedia
Quantity768 units (software)
DistributionNationwide Distribution.