Recall Z-2484-2015 ยท LZG Pump, Infusion, Insulin

Insulet Corporation

OmniPod Insulin Management System. Only the OmniPod device is affected. Catalog Number(s): POD-ZXP420 (US). Intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin.

DateSep 4, 2015 (initiated Jul 13, 2015)
StatusTerminated
ReasonOmniPods (Pods) have a higher rate of failure causing: Cannula fails to deploy/retract causing insulin to be pumped or the audible alarm is displayed on the PDM and pod will not deliver insulin
Root causeNonconforming Material/Component
ActionThe firm, Insulet Corporation, sent an "URGENT: Voluntary Device Recall" letter dated 7/13/2015 to each consignee/customer via Email notification (where possible) on 7/13/15 as well as Federal Express. Customers not responding to the email or Federal Express will receive follow up phone calls. The letter described the product, problem and actions to e taken. The customers were instructed to do the following: If you have Pods from any of these lots, set them aside, we will replace them at no charge; arrange for return and replacement by choosing one of the following methods: 1. Complete the o
Quantity24829 boxes Expanded: 26,230.9 boxes
DistributionWorldwide Distribution: US (Nationwide) and country of: Switzerland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K122953OMNIPOD INSULIN MANAGMENT SYSTEM
Insulet Corporation, Dec 7, 2012