Recall Z-0810-2016 ยท IZI System, X-Ray, Angiographic

Philips Electronics North America Corporation

Actuator for angiographic x-ray monitor ceiling suspension system. Actuator short Assy: THC8AWDS-121 and Actuator long Assy: THC8AWDS-122 used with the following products: 722001 Allura Xper FD 10 C; 722003 Allura Xper FD10; 722005 Allura Xper FD10/10; 722006 Allura Xper FD20; 722008 Allura Xper FD20 Biplane; 722010 Allura Xper FD10; 722011 Allura Xper FD10/10; 722012 Allura Xper FD20; 722013 Al

DateFeb 16, 2016 (initiated Jul 6, 2015)
StatusTerminated
ReasonThe actuator assembly became detached and the monitor carriage with a flexvision monitor dropped to its lowest position.
Root causeProcess control
ActionPhilips sent a Field Safety Notice to all affected customers on July 6, 2015. The notice informs consignees of the reason for recall and actions to be taken by the customer/user. On September 15, 2015, Philips plans to issued a mandatory Field Safety Corrective action (FCO72200280) via Certified Return Receipt. This corrective action is the replacement of the affected MCS actuator. For further information, contact your local representative: Technical Support Line: 1-800-722-9377.
Quantity3158 devices
DistributionWorldwide Distribution - USA (nationwide) including Puerto Rico and the countries of : Canada, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K033737ALLURA XPER FD20
Philips Medical Systems North America, Inc., Dec 9, 2003
K041949ALLURA XPER FD10
Philips Medical Systems North America Co., Jul 30, 2004
K130842ALLURA XPER FD OR TABLE SERIES; ALLURA XPER FD10 SERIES; ALLURA XPER FD20 SERIES; ALLURA XPER FD BIPLANE SERIES
Philips Medical Systems Nederland B.V., May 21, 2013