ClearanceBrief
Recall Z-2087-2015 ยท PGY Display, Diagnostic Radiology

Eizo Corporation

RadiForce GX540, 5MP Monochrome LCD Monitor, Part Code No. 0FTD1826A5, 0FTD1826B5, 0FTD1826C5, UPC 690592034752. This product is intended to be used in displaying and viewing digital images including those of digital mammography for review and analysis by trained medical practitioners.

DateJul 16, 2015 (initiated Jul 7, 2015)
StatusTerminated
ReasonEIZO is recalling the RadiForce LCD Monitors could display abnormally while connecting to particular graphics boards with DisplayPort.
Root causeDevice Design
ActionEIZO sent An Urgent Field Safety Notice dated July 7, 2015, to all customers who purchased the RadiForce RX440, RX650, RX850 and GX540. The notice informs customers that the RadiForce monitors could display abornormally in 6-bit (64 gradations) instead of normal 8-bit (256 gradations) or 10-bit (1024 gradations) while connecting to particular graphics boards with DisplayPort. The notice provides the customers with the actions to be taken and contact information (1-800-800-5202, M-F, 8am-4pm, PST, us.feedback@eizo.com, 5710 Warland Drive, Cypress, CA 90630).
Quantity19 units
DistributionNationwide Distribution including OH, FL, IL, TX, OH, NY, AZ, and NJ.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K1303365MP MONOCHROME LCD MONITOR, RADIFORCE GX540
Eizo Nanao Corporation, Mar 28, 2013