ClearanceBrief
Recall Z-2249-2015 ยท JCT Prosthesis, Tracheal, Expandable

EXP Pharmaceutical Services Corp

Medical Device Exchange 10x40 Airway Stent System; Pouch, Tyvek/Mylar, Sterile, Single Use Only; OEM Catalog number 90129-209; EXP (repackager), Aero (OEM) Prosthesis, tracheal, expandable

DateJul 24, 2015 (initiated Feb 11, 2015)
StatusTerminated
ReasonEXP did not register as a medical device establishment, list the devices being recalled, or establish its own quality system and instead relied on the fact that its vendors were registered and had their own quality systems.
Root causeOther
ActionEXP first notified their customers to quarantine EXP devices February 11-17, 2015. Customers were contacted through telephone, e-mail, and written communications. Customers were instructed to recover and quarantine all EXP products from their inventories. EXP believes that they have contacted all customers to whom products were shipped between April 26, 2014, and December 9, 2014. A second letter is scheduled to be sent in July 2015 (date unknown at this time). This letter will advise users of the recall of due to lack of registration with the FDA.
Quantity1
DistributionDistributed in DC and the states of CO, CA, FL, TX, and AR.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.