Recall Z-2271-2015 ยท BSZ Gas-Machine, Anesthesia
Draeger Medical, Inc.
Fabius MRI Anesthesia Machine; Fabius MRI is an inhalation anesthesia machine for use in MRI environments in operating, induction and recovery rooms.
| Date | Jul 27, 2015 (initiated Jun 22, 2015) |
| Status | Terminated |
| Reason | the Fabius MRI or parts of the system were attracted by the magnetic field of the MRI. If the Fabius MRI is positioned too close to the MRI, its parts can be loosened as a result of the magnetic force of the MRI. |
| Root cause | Labeling False and Misleading |
| Action | Draeger Medical sent an Urgent Medical Device Recall Letter, dated June 2015, and Instructions for Use Supplement (one supplement for each Fabius MRI Anesthesia Machine at the facility) to end users. The letter identified the affected product, problem and actions to be taken. For questions regarding the recall letter contact Michael Kelhart at 1-800-543-5047 (press 1 at the prompt, then 2, then 32349). For questions regarding the operation and/or servicing of the Drger Fabius MRI anesthesia machine contact DrgerService Technical Support at 1-800-543-5047 (press 4 at the prompt). |
| Quantity | 373 |
| Distribution | US Nationwide Distribution |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K072884 | FABIUS MRI Draeger Medical AG & Co. KG, Jul 30, 2008 |