Recall Z-2074-2015 ยท NDP Accessories, Implant, Dental, Endosseous
Nobel Biocare Usa Llc
Abutment Retrieval Instrument Zirconia CC RP/WP, Article No. 37882 and Abutment Retrieval Kit, Article No. 37508 (Article No. 37882 was assembled in this kit). Used to remove a zirconia abutment if the abutment is stuck in the implant connection due to a tight seat. Endosseous dental implant instrument.
| Date | Jul 15, 2015 (initiated Jun 22, 2015) |
| Status | Terminated |
| Reason | One dimension of the affected instrument is incorrect. Therefore this instrument cannot be used to remove the metal adapter as intended. |
| Root cause | Under Investigation by firm |
| Action | A customer notification letter dated 6/22/15 will be sent to all customers who purchased the Abutment Retrieval Instrument Zirconia CC RP/WP because an internal investigation showed that one dimension of the affected instrument is incorrect. Therefore this instrument cannot be used to remove the metal adapter as intended. The letter provides the problems identified and the actions to be taken. Customers are instructed to complete the acknowledgement form and return it to their local Nobel Biocare Subsidiary with the affected instrument. Customers with the affected instrument that is a part |
| Quantity | 149 units total |
| Distribution | Worldwide Distribution-US (nationwide) and the countries of Australia, Austria, Belgium, Canada, Finland, France, Germany, Greece, Hong Kong, Hungary, Italy, Luxembourg, Netherlands, New Zealand, Norway, Portugal, South Africa, Spain, Switzerland, and United Kingdom. |