Recall Z-2266-2015 ยท MWI Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)

Philips Electronics North America Corporation

Invivo Expression MRI Patient Monitoring System.

DateJul 27, 2015 (initiated Jun 17, 2015)
StatusTerminated
ReasonThe device's labeling inaccurately indicates Endotracheal as a body temperature application site. The FlexTEMP reusable temperature sensor is not intended to be used for endotracheal placement. The application site should be listed as Esophageal.
Root causeLabeling False and Misleading
ActionAn URGENT Medical Device Correction was sent to all consignees by letter June 2015 regarding the Invivo Expression MRI Patient Monitoring System FlexTEMP Application Site explaining the reason for the affected labeling.
Quantity697 devices.
DistributionWorldwide Distribution-US (nationwide) including the states of AL, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OR, OK, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY, and Alaska, Hawaii and the District of Columbia, and the countries of
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K124061MRI PATIENT MONITORING SYSTEM (MODEL 865214)
Invivo Corporation, Feb 22, 2013