Recall Z-2102-2015 ยท KNW Instrument, Biopsy

Stryker Instruments Div. of Stryker Corporation

Stryker SelectCore Variable Throw Biopsy Device, 14G, 10CM Product Usage: The Stryker SelectCore Variable Throw Biopsy Device is used in obtaining biopsy samples from soft tissues such as liver, breast, kidney, prostate, spleen, lymph nodes, and various soft tissue tumors. For breast biopsy this product is for diagnosis onlynot therapeutic use.. These devices are sterile, single use devices.

DateJul 18, 2015 (initiated Jun 2, 2015)
StatusTerminated
ReasonStryker Instruments is recalling the Stryker SelectCore Variable Throw Biopsy Device due to the potential for for the stylet to eject from the SelectCore device during actuation.
Root causeEmployee error
ActionStryker distributed Customer Notification Letters via Fed Ex overnight to be delivered on 06/10/2015. Customers are asked to Immediately review this Recall Notification. 2. Immediately check all stock areas and/or operating room storage and quarantine any affected product found. 3. Complete the enclosed Business Reply Form (BRF) to confirm receipt of this notification and identify how many, if any, affected items are currently in your inventory. Please complete and return the BRF even if you do not have any affected product on hand. Note: Your signature on the BRF indicates that you receive
Quantity55 units
DistributionUS Nationwide Distribution in the states of: TX, NJ and MS.