Recall Z-1972-2015 ยท LLZ System, Image Processing, Radiological

Siemens Medical Solutions USA, Inc

SYNGO IMAGING (VERSION V30 and V31); Syngo Imaging is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images, including digital mammography images.

DateJul 2, 2015 (initiated Jun 11, 2015)
StatusTerminated
ReasonDuring certain clinical workflows safety issues may occur involving monitor assignment order and incorrect display and measurements for images with non square pixel size.
Root causeSoftware design
ActionSiemens sent a Customer Advisory Notice, dated June 11, 2015, to their customers. The letter identified the affected product, problem and actions to be taken. Customers were also asked to check the software version and consider whether the issue impacts clinical routine and if a software update is necessary.
Quantity5
DistributionUS in the states of NC and NE
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K071114SYNGO IMAGING (VERSION V30)
Siemens AG Medical Solutions, May 22, 2007
K081734SYNGO IMAGING (VERSION V31)
Siemens AG Medical Solutions, Jul 3, 2008