Recall Z-1933-2015 ยท JAK System, X-Ray, Tomography, Computed

Siemens Medical Solutions USA, Inc

SOMATOM Emotion 16; produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angels or spiral planes taken at different angles.

DateJun 27, 2015 (initiated Jun 4, 2015)
StatusTerminated
ReasonA potential safety issue involving two (2) broken screws and one (1) bowed, which is used to fix the Balance Weight of the DMS in the gantry, was found during a service call at the customer site.
Root causeDevice Design
ActionA safety advisory notice, dated June 4, 2015, was sent to direct accounts explaining the problem and potential risk to patients and operators. Siemens engineers will begin replacing the bolts in May 2015.
Quantity40 total
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133424SOMATOM EMOTION 6 CT SYSTEMS, SOMATOM EMOTION 16
Siemens Medical Solutions USA, Inc., Jan 17, 2014