Recall Z-2785-2015 ยท MUJ System, Planning, Radiation Therapy Treatment

RAYSEARCH LABORATORIES AB

Radiation Therapy Treatment Planning System software RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user.

DateSep 17, 2015 (initiated May 29, 2015)
StatusTerminated
ReasonAn error caused by a floating point precision problem tends to cause a problem with the display of isodose lines for dose and dose difference and color tables for dose, dose difference and PET images
Root causeSoftware design
ActionThe firm, RaySearch Laboratories, distributed a "Field Safety Notice, Medical Device Correction #7106" notice dated May 29, 2015,and reply forms to their customers via email. The notice describes the product, problem and actions to be taken. The Customers were provided a workaround as a temporary fix. Customers were also instructed to educate all users, inspect your product and identify all installed units, confirm you have read and understood the notice by contacting Quality and Regulatory Affairs Manager; and complete and return the reply form via email to freddie.cardel@raysearchlabs.com o
Quantity20 devices
DistributionWorldwide Distribution: US (nationwide) to state of WA; and countries of: France, Sweden, Austria, Spain, Germany, Italy, UK, New Zealand.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K141860RAYSTATION
Raysearch Laboratories AB, Oct 23, 2014