Recall Z-2105-2015 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

ArthroCare Corporation

EVac 70 XTRA with Integrated Cable and PROcise EZ View with Integrated Cable Product Usage: PLASMA WAND devices are indicated for ablation, resection, and coagulation of soft tissue and hemostasis of blood vessels in otorhinolaryngology (ENT) surgery including: adenoidectomy, cysts, head, neck, oral, and sinus surgery, mastoidectomy, myringotomy with effective hemorrhage control, nasal airway obs

DateJul 18, 2015 (initiated Jun 2, 2015)
StatusTerminated
ReasonWands inspected with the adaptor fixture were released to the field not meeting the dielectric (Hipot) specifications.
Root causeNonconforming Material/Component
ActionSmith&Nephew sent an Urgent Notification letter dated June 10, 2015 via mail/courier service with delivery tracking/confirmation to affected customers. A response form is included in the notification letter. A follow up letter will be sent to non-responding consignees. Customers were instructed to complete the acknowledgement form and return via fax or email.
Quantity733 devices
DistributionWorldwide Distribution - US Nationwide and the countries of Australia, Belgium, Germany, Spain, United Kingdom, Ireland, India, Italy and Singapore.
FDA recordOpen on fda.gov