Recall Z-1989-2015 ยท KWQ Appliance, Fixation, Spinal Intervertebral Body

Alliance Partners LLC

Nakoma-SL Anterior Cervical Plating System's Variable Single Barrel Drill Guide Product Usage: The Nakoma-SL ACP System is intended for anterior intervertebral screw fixation of the cervical spine at levels C2-T1. The Nakoma ACP System is indicated for temporary stabilization of the anterior spine during the development of cervical spine fusion in patients with the following indications: Degenera

DateJul 7, 2015 (initiated Jun 9, 2015)
StatusTerminated
ReasonInadequate interference between the screw and the plate when screw implanted into the plate at angulation greater than 7 degrees can result in the screw advancing through the plate during implantation.
Root causeDevice Design
ActionAlliance Spine sent an Urgent Medical Device Field Action Notification letter dated June 15, 2015 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete the response form and fax to 210-314-2524.
Quantity21 units
DistributionUS Nationwide Distribution in the states LA, TX, and PR.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K141993NAKOMA ACP
Alliance Partners, LLC, Nov 6, 2014