Recall Z-1849-2015 ยท JAQ System, Applicator, Radionuclide, Remote-Controlled

Varian Medical Systems Inc

The VariSource iX series afterloader systems are computer controlled remote electro/mechanical systems used for medical purposes, for placing a cable incorporating an irradiated iridium seed internally or close by a malignant tumor or tumor bed in a practice known as brachytherapy.

DateJun 26, 2015 (initiated May 18, 2015)
StatusTerminated
ReasonAfter an application freeze and restart of the VariSource iX series by power cycling, the Partial Fraction generated by the system will not be correct. The application freeze issue affects only the systems equipped with magnetic Hard Disk Drives (HDD).
Root causeSoftware Design Change
ActionAll the consignees affected by this recall will be notified by: The Urgent Medical Device Correction Letter/ Field Safety Notification will be distributed to all affected users, with a description of the problem. The Letter will also be distributed to the Varian Sales, Marketing, and Service organizations, informing them of the issue. Additionally, the letter will be posted to the customer support site: http:/ /www.MyVarian.com.
Quantity55
DistributionWorldwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K122236VARISOURSE IX BRAHYTHERAPY AFTERLOADER VARISOURCE IX (T) BRACHYTHERAPY AFTERLOADER
Varian Medical Systems, Inc., Nov 16, 2012