Recall Z-1842-2015 ยท MDB System, Blood Culturing
Remel Inc
VersaTREK Instrument Series 528 REF 6528, Manufactured for Remel Inc. The VersaTREK System is for cultivating and recovering microorganisms, especially bacteria and yeasts, from blood and other normally sterile bodily fluids.
| Date | Jun 25, 2015 (initiated May 19, 2015) |
| Status | Terminated |
| Reason | Use of the recalled product may result in false positive reports. |
| Root cause | Release of Material/Component prior to receiving test results |
| Action | Remel, Inc. a part of Thermo Scientific sent a Product Correction Notification dated May 19, 2015, to all affected customers. The notification states the problem with the device and that the recalling firm's Technical Services department will be contacting the customer to schedule a field service of the device to correct the issue. Customers were instructed to notify any personnel who need to be notified of the potential for a false positive resulting from rare inconsistent barometric pressure readings. Return the attached Medical Device Recall Return Response form acknowledging receipt of |
| Quantity | 21 devices |
| Distribution | Worldwide distribution: US (nationwide) to states of AZ, CO, FL, GA, IA, ID, LA, MI, MO, PR, TN, TX, VA, and WA and the countries of Canada, People's Republic of China, Egypt, Finland, France, Indonesia, Israel, Jordan, Kenya, Republic of Korea, Latvia, Romania, Russia, Spain, Thailand, United Kingd |