Recall Z-2043-2015 ยท IYO System, Imaging, Pulsed Echo, Ultrasonic

Volcano Corp

Volcano s5/s5x/CORE Mobile Intravascular Ultrasound Imaging and Pressure System; Product/Part numbers: s5 - 804200-002 and 804200-014; s5x - 807300-001; CORE MOBILE - 400-0100.01. Intravascular Imaging and Pressure System for evaluation of blood flow, perfusion and vascular structures.

DateJul 8, 2015 (initiated Sep 23, 2013)
StatusTerminated
ReasonA limitation in interoperability between the Philips Xper Flex Cardia Physiomonitoring System and Volcano's sS/Si, Core, and ComboMap systems. This limitation may impact the calculation of Fraction Flow Reserve (FFR) within the Volcano System when the system is connected to the Xper Flex Cardia system via its aortic pressure auxiliary output.
Root causeNonconforming Material/Component
ActionVolcano and Philips sent an Advisory Notice letter dated September 23, 2013 to all impacted customers. The letter identified the affected product, problem and actions to be taken. For question contact your Volcano and Philips representative.
Quantity227 total units affected by the issue - all models
DistributionWorldwide Distribution - US Nationwide and the countries of Belgium, United Kingdom, Denmark, Netherlands, Sweden, Iceland, France, Saudi Arabia, Australia, Philippines and Thailand.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K123898VOLCANO S5/S5I INTRAVASCULAR ULTRASOUND IMAGING AND PRESSURE SYSTEM
Volcano Corporation, Jan 14, 2013