ClearanceBrief
Recall Z-1787-2015 ยท KPS System, Tomography, Computed, Emission

Philips Medical Systems (Cleveland) Inc

GEMINI TF PET/CT 64 Slice (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.

DateJun 11, 2015 (initiated Mar 6, 2015)
StatusTerminated
ReasonThe firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.
Root causeNonconforming Material/Component
ActionOn 3/9/2015 the firm sent Urgent Field Safety Notices Medical Device Correction letters to their customers.
Quantity71 units
DistributionWorldwide Distribution-US (nationwide) including DC & Puerto Rico and the following countries: Albania, Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia, Brazil, Bulgaria, Cambodia, Cameroon, Cayman Islands, China, Colombia, Democratic Congo, Croatia, Cuba, Czech Republic, Denmar
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052640GEMINI RAPTOR
Philips Medical Systems (Cleveland), Inc., Oct 7, 2005