Recall Z-1776-2015 ยท KPS System, Tomography, Computed, Emission
Philips Medical Systems (Cleveland) Inc
Gemini 16 PET/CT System (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.
| Date | Jun 11, 2015 (initiated Mar 6, 2015) |
| Status | Terminated |
| Reason | The firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes. |
| Root cause | Nonconforming Material/Component |
| Action | On 3/9/2015 the firm sent Urgent Field Safety Notices Medical Device Correction letters to their customers. |
| Quantity | 7 units |
| Distribution | Worldwide Distribution-US (nationwide) including DC & Puerto Rico and the following countries: Albania, Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia, Brazil, Bulgaria, Cambodia, Cameroon, Cayman Islands, China, Colombia, Democratic Congo, Croatia, Cuba, Czech Republic, Denmar |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K041955 | GEMINI PET/CT IMAGING SYSTEM Philips Medical Systems (Cleveland), Inc., Jul 29, 2004 |