Recall Z-1971-2015 ยท KHO Fluorometer, For Clinical Use

Tecan US, Inc.

Infinite 200 in vitro diagnostic. Product Usage: The Tecan Infinite 200 is a multifunctional microplate reader with injector option. The Infinite 200 provides high performance for the vast majority of todays microplate applications and research and is robotic compatible. The Infinite 200 PRO has been designed as a general purpose laboratory instrument for professional use, supporting common 6

DateJul 2, 2015 (initiated May 29, 2015)
StatusTerminated
ReasonA firmware bug causes incorrect values to be displayed for specific measured luminescence values.
Root causeDevice Design
ActionTecan sent an Urgent Field Corrective Action letter dated May 29, 2015 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete the attached Customer Response Form. For questions contact your local Tecan Helpdesk.
Quantity868 in total