Recall Z-1969-2015 ยท KHO Fluorometer, For Clinical Use
Tecan US, Inc.
Infinite M1000 in vitro diagnostic. Product Usage: The Tecan Infinite M1000 is a multifunctional monochromator based microplate reader that provides high performance for the vast majority of todays microplate applications and research. Besides offering absorbance and fluorescence intensity measurements, the Infinite M1000 allows the measurement of fluorescence polarization and luminescence. The
| Date | Jul 2, 2015 (initiated May 29, 2015) |
| Status | Terminated |
| Reason | A firmware bug causes incorrect values to be displayed for specific measured luminescence values. |
| Root cause | Device Design |
| Action | Tecan sent an Urgent Field Corrective Action letter dated May 29, 2015 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete the attached Customer Response Form. For questions contact your local Tecan Helpdesk. |
| Quantity | 868 in total |
| Distribution | US Nationwide Distribution in the states of AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MO, NE NH, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, including DC |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.