Recall Z-1968-2015 ยท KHO Fluorometer, For Clinical Use

Tecan US, Inc.

Infinite M1000 PRO in vitro diagnostic. Product Usage: The Tecan Infinite M1000 PRO is a multifunctional monochromator based microplate reader that provides high performance for the vast majority of todays microplate applications and research. In addition to offering absorbance and fluorescence intensity measurements, the Infinite M1000 PRO can also perform fluorescence polarization and lumines

DateJul 2, 2015 (initiated May 29, 2015)
StatusTerminated
ReasonA firmware bug causes incorrect values to be displayed for specific measured luminescence values.
Root causeDevice Design
ActionTecan sent an Urgent Field Corrective Action letter dated May 29, 2015 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete the attached Customer Response Form. For questions contact your local Tecan Helpdesk.
Quantity868 in total
DistributionUS Nationwide Distribution in the states of AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MO, NE NH, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, including DC