Recall Z-1967-2015 ยท KHO Fluorometer, For Clinical Use
Tecan US, Inc.
Infinite F500, in vitro diagnostic. Product Usage: The Tecan Infinite F500 is a multifunctional microplate reader with injector option. The Infinite F500 provides high performance for the vast majority of todays microplate applications and research and is robotic compatible. The Infinite F500 is intended as a general purpose laboratory instrument for professional use, supporting common micropla
| Date | Jul 2, 2015 (initiated May 29, 2015) |
| Status | Terminated |
| Reason | A firmware bug causes incorrect values to be displayed for specific measured luminescence values. |
| Root cause | Device Design |
| Action | Tecan sent an Urgent Field Corrective Action letter dated May 29, 2015 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete the attached Customer Response Form. For questions contact your local Tecan Helpdesk. |
| Quantity | 868 in total |
| Distribution | US Nationwide Distribution in the states of AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MO, NE NH, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, including DC |
| FDA record | Open on fda.gov |