Recall Z-2207-2015 ยท MMI Immunoassay Method, Troponin Subunit

Siemens Healthcare Diagnostics, Inc

ADVIA Centaur Tni-Ultra Assay; SMN 10317708 - 100 tests and SMN 10317709 -500 tests Product Usage: For in vitro diagnostic use in the quantitative determination of cardiac troponin I (cTnl) in serum, heparinized plasma, or EDTA plasma using the ADVIA Centaur/ADVIA Centaur XP/ADVIA Centaur XPT/ADVIA Centaur CP systems. Cardiac troponin I determinations aid in the diagnosis of acute myocardial infa

DateJul 23, 2015 (initiated May 19, 2015)
StatusTerminated
ReasonSystem-to-system bias between the Tni-Ultra assay on the ADVIA Centaur CP and ADVIA Centaur/ADVIA Centaur XP/ADVIA Centaur XPT systems.
Root causeUnder Investigation by firm
ActionSiemens sent an Urgent Medical Device Correction letter dated via Federal Express to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete and return the Field Correction Effectiveness Check attached to the letter. For questions contact your Siemens Customer Care Center or your local Siemens technical support representative.
QuantitySMN10317708 = 20544 kits & SMN 10317709 = 21159 kits
DistributionWorldwide distribution - US Nationwide in the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MO, MN, MS, MT, NC, NJ, NM, NY, NV, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WY, WV, and Puerto Rico; and the countries of Algeria, Austria, Bahrain, Belgium
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K053020TNL-ULTRA ASSAY FOR THE ADVIA CENTAUR SYSTEM
Bayer Healthcare, LLC, Dec 30, 2005