ClearanceBrief
Recall Z-1739-2015 ยท KWZ Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer

Smith & Nephew, Inc.

R3(TM) 28 MM I.D., 62 MM O.D. CONSTRAINED LINER, REF 71339162, QTY: (1), STERILE EO. Product Usage: The R3 Constrained liner Acetabular System is a cemented or uncemented prosthesis intended to replace a hip joint. The Constrained liner is intended for primary or revision patients at high risk for hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular

DateJun 9, 2015 (initiated May 6, 2015)
StatusTerminated
ReasonThe R3 Constrained outer liner locking groove diameter is 0.030 inches oversized and may not allow the locking ring to engage into the R3 shell.
Root causeProcess change control
ActionSmith & Nephew sent an Urgent-Product Recall Notification letter dated May 6, 2015. to Sales offices by E-mail and Hospitals via Fed- Ex. The letter identified the affected product, problem and actions to be taken. Customers were asked to quarantine the product and return it. Distributors were asked to notified their customers of the recall and ask them to return the product.
Quantity32 units
DistributionWorldwide Distribution - US Nationwide in the states of CA, FL, IL, IN, MI, NE, NY, TN, TX, VA, WI and the countries of Australia, Switzerland, Chile, Canada, and Spain
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K083566R3 CONSTRAINED LINER, CONSTRAINED LINER
Smith & Nephew, Inc., Mar 3, 2009