Recall Z-1737-2015 ยท KWZ Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer
Smith & Nephew, Inc.
R3(TM) 22 MM I.D., 58 MM O.D. CONSTRAINED LINER, REF 71339158, QTY: (1), STERILE EO. Product Usage: The R3 Constrained liner Acetabular System is a cemented or uncemented prosthesis intended to replace a hip joint. The Constrained liner is intended for primary or revision patients at high risk for hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular
| Date | Jun 9, 2015 (initiated May 6, 2015) |
| Status | Terminated |
| Reason | The R3 Constrained outer liner locking groove diameter is 0.030 inches oversized and may not allow the locking ring to engage into the R3 shell. |
| Root cause | Process change control |
| Action | Smith & Nephew sent an Urgent-Product Recall Notification letter dated May 6, 2015. to Sales offices by E-mail and Hospitals via Fed- Ex. The letter identified the affected product, problem and actions to be taken. Customers were asked to quarantine the product and return it. Distributors were asked to notified their customers of the recall and ask them to return the product. |
| Quantity | 12 units |
| Distribution | Worldwide Distribution - US Nationwide in the states of CA, FL, IL, IN, MI, NE, NY, TN, TX, VA, WI and the countries of Australia, Switzerland, Chile, Canada, and Spain |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K083566 | R3 CONSTRAINED LINER, CONSTRAINED LINER Smith & Nephew, Inc., Mar 3, 2009 |