Recall Z-1832-2015 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

ArthroCare Medical Corporation

EVAC 70 XTRA with Integrated Cable REF EICA5872-01 Product Usage: PLASMA WAND devices are indicated for ablation, resection, and coagulation of soft tissue and hemostasis of blood vessels in otorhinolaryngology (ENT) surgery including: adenoidectomy, cysts, head, neck, oral, and sinus surgery, mastoidectomy, myringotomy with effective hemorrhage control, nasal airway obstruction by reduction of h

DateJun 25, 2015 (initiated May 15, 2015)
StatusTerminated
ReasonPotential component failure causing device inoperability
Root causeComponent design/selection
ActionSmith & Nephew sent an Urgent Product Recall notification letter dated May 15, 2015 to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to inspect inventory, quarantine immediately and return to the address listed on the inventory Return Certification Form.
Quantity4,111 units
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K142999EVac 70 Xtra Plasma Wand with Integrated Cable, PROcise XP Plasma Wand with Integrated Cable
ArthroCare Corporation, Nov 19, 2014