Recall Z-1932-2015 ยท CEM Electrode, Ion Specific, Potassium

Abbott Laboratories

ICT Serum Calibrator REF 146-03. For use in the calibration of the serum Sodium, Potassium, and Chloride assays.

DateJun 27, 2015 (initiated Jul 22, 2014)
StatusTerminated
ReasonICT Serum Calibrator may generate lower than expected Potassium Quality Control (QC) and patient results on the Abbott Architect cSystem.
Root causeProcess control
ActionCustomers who received shipments of the affected lots were sent a Product Correction letter on July 22, 2014.
Quantity11,215 units
DistributionWorldwide Distribution-US (nationwide) including Puerto Rico, and the countries of Afghanistan, Algeria, Andorra, Angola, Argentina, Australia, Austria, Bahamas, Bangladesh, Belarus, Belgium, Bermuda, Bosnia & Herzegovina, Brazil, Brunei, Bulgaria, Burkina Faso, Canada, Cayman Islands, Chile, China,