Recall Z-1638-2015 ยท OXO Image-Intensified Fluoroscopic X-Ray System, Mobile

GE OEC Medical Systems, Inc

GE OEC Brivo 865 plus, Mobile X-ray system laser aimer assemblies (part number 520002DW)

DateJun 8, 2015 (initiated May 22, 2014)
StatusTerminated
ReasonLaser aimer assemblies were shipped without certification of the overall assembly or submitting a laser product report and therefore do not comply with the US regulations for laser products.
Root causeRadiation Control for Health and Safety Act
ActionGE Healthcare Surgery Planned Action(s) to repair the defect or to bring the product into compliance: The Center for Devices and Radiological Health (CDRH) understands that the CAP involving 9 units consisted of the following: 1. Replacing the tube side cover with the laser aimer assembly with a tube side cover without a laser aimer assembly. This correction brought the systems into compliance with applicable US regulations. For further questions please call (801) 536-4615.
Quantity9
DistributionUS Distribution to the states of: MD, NC, MT, NJ, UT and NY
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K123603OEC BRIVO 865 ADVANCE, OEC BRIVO 785 ESSENTIAL, AND OEC BRIVO 715 PRIME
Ge Healthcare Surgery GE Oec Medical Systems, Inc., May 17, 2013