Recall Z-1719-2015 ยท IWB System, Radiation Therapy, Radionuclide

Elekta, Inc.

Laksell GammaPlan, a computer based dose planning system specifically designed for use with Leksell Gamma Knife, radiation therapy treatment.

DateJun 5, 2015 (initiated Jun 1, 2015)
StatusTerminated
ReasonMemory can become corrupted when creating a fused study via drag and drop in Leksell GammaPlan 10.2.
Root causeSoftware in the Use Environment
ActionThe firm, Elekta, sent an "URGENT IMPORTANT FIELD SAFETY NOTICE 100-01-102-015" to all customers on 6/1/2015. The notice described the product, problem and actions to be taken. The customers were instructed to not use drag-and-drop to fuse images as a solution until a permanent fix can be developed and implemented on all affected devices. A solution is in development that will resolve the issue and is expected to be released September 2015. Elekta Service will then work with customers to implement on affected devices - estimated 6 month completion. Customers were also instructed to complete
Quantity10
DistributionWorldwide Distribution: US (nationwide) including states: IL, OH, PA, TX, and WI; and countries of: Germany, Japan and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K103093LEKSELL GAMMAPLAN
Elekta Instrument AB, Dec 21, 2010