Recall Z-1646-2015 ยท JAK System, X-Ray, Tomography, Computed

Siemens Medical Solutions USA, Inc

SOMATOM Perspective; the intended use of computed tomography is to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angels or spiral planes taken at different angles.

DateMay 22, 2015 (initiated Apr 20, 2015)
StatusTerminated
ReasonSoftware bugs in VC20B SP0a or SP1 software versions may cause issues that could make it necessary to rescan patients. Syngo Main UI may crash if the Patient Browser is scrolled with arrow keys of the keyboard. Tomo images may freeze on the screen when zoomed in or out under the mode of CAREVision. Sporadic displaying error in WorkStream4D application. DB may lock under heavy and multitasks.
Root causeSoftware design
ActionSiemens sent an Customer Safety Advisory Notice dated April 20, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. A remote software update is scheduled for release that will resolve the issue. For further questions, please call (610) 219-6300.
Quantity75
DistributionNationwide Distribution including Puerto Rico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133590SOMATOM PERSPECTIVE
Siemens Medical Solutions USA, Inc., Apr 30, 2014