ClearanceBrief
Recall Z-1841-2015 ยท MDB System, Blood Culturing

Remel Inc

VersaTREK Instrument Series 240 REF 6240, Manufactured for Remel Inc. The VersaTREK System is for cultivating and recovering microorganisms, especially bacteria and yeasts, from blood and other normally sterile bodily fluids.

DateJun 25, 2015 (initiated May 19, 2015)
StatusTerminated
ReasonUse of the recalled product may result in false positive reports.
Root causeRelease of Material/Component prior to receiving test results
ActionRemel, Inc. a part of Thermo Scientific sent a Product Correction Notification dated May 19, 2015, to all affected customers. The notification states the problem with the device and that the recalling firm's Technical Services department will be contacting the customer to schedule a field service of the device to correct the issue. Customers were instructed to notify any personnel who need to be notified of the potential for a false positive resulting from rare inconsistent barometric pressure readings. Return the attached Medical Device Recall Return Response form acknowledging receipt of
Quantity143 devices
DistributionWorldwide distribution: US (nationwide) to states of AZ, CO, FL, GA, IA, ID, LA, MI, MO, PR, TN, TX, VA, and WA and the countries of Canada, People's Republic of China, Egypt, Finland, France, Indonesia, Israel, Jordan, Kenya, Republic of Korea, Latvia, Romania, Russia, Spain, Thailand, United Kingd
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.