Recall Z-1958-2015 ยท OWB Interventional Fluoroscopic X-Ray System

Siemens Medical Solutions USA, Inc

Cios Alpha; The Cios Alpha is a mobile X-ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to: interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patien

DateJul 2, 2015 (initiated May 5, 2015)
StatusTerminated
Reasonunder certain circumstances the Cios Alpha system may freeze during a procedure.
Root causeSoftware design
ActionSiemens sent an Safety Advisory Notice dated May 5 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. They were alerting them about the issue and risks involved. Customers were provided work-around instructions until a software fix is released. For further questions please call (610) 219-6300.
Quantity37
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K132094CIOS ALPHA
Siemens Medi Cal Solutions, Inc., Mar 11, 2014