Recall Z-1734-2015 ยท LZG Pump, Infusion, Insulin

Roche Diabetes Care, Inc.

AccuChek Spirit insulin infusion pump.

DateJun 9, 2015 (initiated Apr 22, 2015)
StatusTerminated
ReasonRoche has become aware of an increase of mechanical errors experienced by customers using the Accu-Chek Spirit Combo insulin infusion pump and Accu-Chek Spirit insulin infusion pump when the insulin pumps display E6 and El0 error messages. If the user does not act upon the E6 and E10 error messages appropriately, insulin delivery will be stopped and if unnoticed may lead to severe hyperglycemia u
Root causeDevice Design
ActionRoche sent an URGENT MEDICAL DEVICE CORRECTION notification dated April 22, 2015, to all affected customers. The letter identified the problem the product and the action needed to be taken by the customer. The recall notification included a description of the reason for the recall, customer responsibilities, training chart, and instructions for responding to the formal recall notification. Please contact the ACCU-CHEK Customer Care Service Center, 24 hours a day, seven days a week at 1-800-688-4578, if you have questions about the information contained in this notification.
Quantity35,795
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K042887ACCU-CHEK SPIRIT
Disetronic Medical Systems AG, Mar 18, 2005