ClearanceBrief
Recall Z-1608-2015 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Spacelabs Healthcare Inc

Spacelabs qube Compact Monitor, Model 91390. The monitor is functioning as either bedside or central monitors; passively displays data generated by Spacelabs Healthcare parameter modules, Flexports interfaces, and other SDLC based products in the form of waveform and numeric displays, trends and alarms.

DateMay 13, 2015 (initiated May 8, 2015)
StatusTerminated
ReasonThe input power can fail even though the power source at the wall has not been interrupted. In addition, there is the potential for the qube Compact Monitor to fail to turn on. This is a retrospective recall related to a recall initiated in 2012 (Z-0403-2013).
Root causeSoftware design
ActionThe firm sent the Urgent Medical Device Correction- Spacelabs Healthcare Inc.-qube Compact Monitor, Model 91390, Power Input PCBA Failure letter, dated 8 May 2015, via priority service mail to US consignees. This letter described the affected system, problem, actions to be taken by users and corrective actions conducted by Spacelabs. For additional information or technical assistance, please contact: Technical Support Spacelabs Healthcare, Inc. 35301 SE Center St. Snoqualmie, WA 98065 1-800-522-7025 and select 2 for Technical Support
Quantity5 units distributed in the US
Distributiondistributed in North Carolina, Texas, and Wyoming
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K120616SPACELABS HEALTHCARE QUBE COMPACT MONITOR (91390
Spacelabs Medical, Inc., Mar 29, 2012