Recall Z-1695-2015 ยท JAK System, X-Ray, Tomography, Computed
Philips Medical Systems (Cleveland) Inc
Brilliance iCT Computed Tomography X-ray system
| Date | May 29, 2015 (initiated Apr 13, 2015) |
| Status | Terminated |
| Reason | A calibrated torque wrench malfunctioned at the supplier leading to possible incorrect torque of x-ray tube mounting fasteners. |
| Root cause | Other |
| Action | Philips sent an Urgent - Field Safety Notice dated April 13, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Philips Healthcare is recommending to stop the use of your system until the X-ray tube mounting fasteners are replaced. In addition, a Philips Field Service Engineer (FSE) will service your system via a Mandatory Field Change Order (FCO 72800638). A Philips FSE will contact you to schedule this appointment. If you need any further information or support concerning this issue, please contact your local |
| Quantity | 2 |
| Distribution | Internationally to the countries of China and Norway. |
| FDA record | Open on fda.gov |