ClearanceBrief
Recall Z-1925-2015 ยท LBS Ldl & Vldl Precipitation, Cholesterol Via Esterase-Oxidase, Hdl

Roche Diagnostics Operations, Inc.

HDL-Cholesterol plus gen.3.; and Small Supersized Intended for use on COBAS Integra 800 and 400 plus systems, cobas c111, and cobas c systems for the quantitative determination of HDL-cholesterol concentration in serum and plasma.

DateJun 27, 2015 (initiated May 7, 2015)
StatusTerminated
ReasonPossible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.
Root causeUnder Investigation by firm
ActionRoche Diagnostics initiated a voluntary recall and the Urgent Medical Device Correction notices were sent via UPS Ground on 05/07/2015 to customers. Customers were notified of the interference and the method sheets and package inserts will be updated with the new interference claims. The firm plans to monitor any non-responding accounts and address those customers with follow-up attempts to distribute recall notices. The firm plans to conduct level B effectiveness checks via fax and phone. Until the new method sheets are available, the Urgent Medical Device Correction serves as the updated
Quantity6255
DistributionUS (nationwide) including PR.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K012286MODIFICATION TO:COBAS INTEGRA HDL-CHOLESTEROL PLUS 2ND GENERATION
Roche Diagnostics Corp., Aug 8, 2001