ClearanceBrief
Recall Z-1922-2015 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Roche Diagnostics Operations, Inc.

Creatinine Plus. Intended for use on Hitachi/MODULAR systems for the quantitative determination of the creatinine concentration in serum, plasma, and urine.

DateJun 27, 2015 (initiated May 7, 2015)
StatusTerminated
ReasonPossible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.
Root causeUnder Investigation by firm
ActionRoche Diagnostics initiated a voluntary recall and the Urgent Medical Device Correction notices were sent via UPS Ground on 05/07/2015 to customers. Customers were notified of the interference and the method sheets and package inserts will be updated with the new interference claims. The firm plans to monitor any non-responding accounts and address those customers with follow-up attempts to distribute recall notices. The firm plans to conduct level B effectiveness checks via fax and phone. Until the new method sheets are available, the Urgent Medical Device Correction serves as the updated
Quantity6,255 analytical units distributed
DistributionUS (nationwide) including PR.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K953239BOEHRINGER MANNHEIM/HITACHI 917 ANALYZER
Boehringer Mannheim Corp., Sep 25, 1995