Recall Z-1606-2015 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

GE Healthcare

GE Healthcare, CARESCAPE Monitor B850, B650 and B450 (Bx50).

DateMay 12, 2015 (initiated May 4, 2015)
StatusTerminated
ReasonThe heart rate could be calculated from pacer pulses without indication that pacemaker detection is OFF in some combinations of Bx50 monitors, a PDM, and CIC/CSCS. Then the monitor does not show the Pacer Off message and there may not be an alarm for asystole. Undetected asystole could result in irreversible changes in the patients condition and delayed or missed life sustaining patient treatment
Root causeSoftware design
ActionGE Healthcare sent an Urgent Medical Device Correction letter dated May 1, 2015, to all affected consignees. The letter was addressed to Healthcare Administrator / Risk Manager, Chief of Nursing, and Director of Biomedical Engineering. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Quantity110 (83 units US, 27 units OUS)
DistributionWorldwide Distribution - US including CO, MI, WI and Internationally to NETHERLANDS, FINLAND, and BELGIUM.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K092027CARESCAPE MONITOR B850
Ge Medical Systems Information Technologies, Aug 28, 2009
K102239CARESCAPE MONITOR B650
GE Healthcare Finland Oy, Oct 18, 2010
K131223CARESCAPE MONITOR B650
GE Healthcare Finland Oy, Aug 28, 2013
K131414CARESCAPE MONITOR B850
GE Healthcare Finland Oy, Aug 28, 2013
K132533CARESCAPE MONITOR B450
GE Healthcare Finland Oy, Dec 4, 2013