Recall Z-1642-2015 ยท GCJ Laparoscope, General & Plastic Surgery

DeRoyal Industries Inc

DeRoyal(R) Medium Specimen Retrieval Bag, REF 5500-000-003, Rx Only Product Usage: The device is a single-use specimen retrieval bag indicated for use in laparoscopic procedures to capture and remove organs or tissue from the body cavity.

DateMay 21, 2015 (initiated Apr 17, 2015)
StatusTerminated
ReasonThe firm received reports of specimen retrieval bags tearing, ripping, and coming apart during use.
Root causeDevice Design
ActionDeRoyal sent an Urgent: Voluntary Recall letter dated April 17, 2015 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to 1. Check your inventory for the product and specific lot number from the attached Recall Response Form. 2. If you have affected inventory, please destroy it and dispose of it according to your facilitys procedures. 3. Record all information on the included Recall Response and Notice of Destruction forms. 4. If you purchase through a distributor, please indicate their name/location on the Notice of D
Quantity5359 units
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K132375GENISTRONG SINGLE-USE SPECIMEN RETRIEVAL BAG
Genicon, Jun 12, 2014