Recall Z-1607-2015 ยท LMC Camera, Multi Format, Radiological

Philips Medical Systems (Cleveland) Inc

Nuclear gamma cameras (ADAC VERTEX PLUS/SOLUS/CARDIO, ADAC VERTEX V60 & ADAC CARDIO C60) Nuclear gamma cameras intended to produce images depicting the anatomical distribution of single photon and positron emitting radioisotopes with the human body for interpretation by medical personnel.

DateMay 13, 2015 (initiated Mar 11, 2015)
StatusTerminated
ReasonThe firm was notified by a customer that the collimator exchange carriage and the collimator storage cabinet were misaligned.
Root causeDevice Design
ActionPhilips sent an Urgent - Medical Device Correction letter dated March 12, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The firm requested their customers to do the following: " If a misalignment on the top and/or bottom of the guide rail between the cabinet and carriage can be visually detected do not continue with the collimator exchange procedure, " If the misalignment is visually detected, be aware that the customer/user cannot remedy this issue and will require service interaction prior to the continued
Quantity975 units
DistributionWorldwide Distribution - US (nationwide) to the states of: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and WY., and to the countries of : Australia, Austria