Recall Z-1605-2015 ยท DZE Implant, Endosseous, Root-Form

BioHorizons Implant Systems Inc

Tapered HD Upgrade Set, REF TSKHDUS, Non-Sterile, Rx Only, Product Usage Usage: Surgical instruments used in preparation for dental implant placement

DateMay 12, 2015 (initiated Apr 16, 2015)
StatusTerminated
ReasonBioHorizons Tapered HD Upgrade set was packaged with an incorrect drill component manufactured of stainless steel in place of the correct component manufactured from titanium.
Root causeProcess control
ActionBioHorizons sent an Urgent Medical Device Field Correction letter on 04/16/2015 to affected customers. Non-responding customers will be notified after 30 days. The letter identified the affected product, problem and actions to be taken. For questions contact Customer Care Department at 888-246-8338.
Quantity48 units
DistributionUS Nationwide in the states of: AL, FL, MS, VA, LA, CA, NC, TX, TN, OH, OK, VT, WA, MD, NY, SD, and SC.