Recall Z-1811-2015 ยท FRN Pump, Infusion

CareFusion 303, Inc.

Alaris PC unit model 8015, Infusion pump. software version 9.17 Product Usage: The Alaris PC unit is the central programming, monitoring and power supply component for the Alaris System. The software is embedded in the Alaris PC unit.

DateJun 16, 2015 (initiated May 13, 2015)
StatusTerminated
ReasonAn issue with the cancel functionality used during atypical infusion programming to cancel user inputted values. An infusion may start that is greater than or less than the hospital established limits for the specific medication.
Root causeSoftware design
ActionCareFusion sent an Urgent Medical Device Recall Notification letter dated May 13, 2015 to customers. The letter identified the affected product, problem and the recommended actions to be taken. Customers are informed that CareFusion has released a software upgrade to address this issue. CareFusion will contact all affected customers within 60 days to initiate the scheduling process for the software upgrade installation. Customers with recall related questions are instructed to contact CareFusion Support Center at (888) 562-6018, 7am-4pm PT, supportcenter@carefusion.com. Customers with ad
Quantity63 units
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051641ALARIS SYSTEM PC UNIT, MODEL 8001
Cardinal Health, Alaris Products, Oct 20, 2005