Recall Z-1634-2015 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Stryker Howmedica Osteonics Corp.

Stryker Orthopaedics Scorpio femoral component. Howmedica Osteonics Corp., A subsidiary of Stryker Corporation Made in USA Sterile. For use in total reconstructive replacement of the knee joint.

DateMay 19, 2015 (initiated Aug 15, 2014)
StatusTerminated
ReasonThe peel strength of the inner blister may have been below internal validated requirements. The strength of the seal is directly related to long term shelf life of the sterile barrier.
Root causeUnder Investigation by firm
ActionThe firm, Stryker Orthopaedics, notified their Branches/Agencies via email on 8/15/2014 and sent an "Urgent Product Recall" letter dated 8/15/2014 with a Product Recall Acknowledgement Forms to their Branches/Agencies and consignees. The letter describes the product, issue, potential hazards and actions to be taken. The consignees were instructed to complete and return the attached Product Recall Acknowledgment Form within 5 days via fax to 855-251-3635; and return the affected product to the attention of Regulatory Compliance, Stryker Orthopaedics, 325 Corporate Drive, Mahwah NJ 07430. If
Quantity1,147 in total
DistributionWorldwide Distribution: US (nationwide) including PR and countries of: Australia, Canada, China, Brazil, France, Hong Kong, India, Italy, Mexico, Netherlands, Singapore, South Africa, Sweden, and Spain.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K030978SCORPIO NRG KNEE SYSTEM
Howmedica Osteonics Corp., Nov 20, 2003
K042343SCORPIO NRG KNEE SYSTEM
Howmedica Osteonics Corp., Feb 9, 2005
K962152OSTEONICS SCORPIO POSTERIORLY STABILIZED TOTAL KNEE SYSTEM
Osteonics Corp., Aug 14, 1996
K974556OSTEONICS SCORPIO POSTERIOR CRUCIATE RETAINING TOTAL KNEE SYSTEM
Osteonics Corp., Feb 27, 1998