Recall Z-1706-2015 ยท BSZ Gas-Machine, Anesthesia

Maquet Cardiovascular Us Sales, Llc

FLOW-i Anesthesia System The FLOW-i Anesthesia System is intended for use administering inhalation anesthesia on neonatal to adult patient populations.

DateJun 8, 2015 (initiated Apr 2, 2015)
StatusTerminated
ReasonIn some Flow-i Anesthesia Systems, patient cassette can get dislodged which could result in gas leakage within the system internal circuit potentially leading to a stop in ventilation.
Root causeUnder Investigation by firm
ActionThe firm, Maquet Inc., sent an "Urgent Medical Device Field Action Field Correction for FLOW-i Anesthesia Systems" letter dated 4/2/2015 via FedEx. The letter describes the product, problem and actions to be taken. The customers were instructed to read the letter and inform staff, who may use the FLOW-i Anesthesia System of the patient cassette locking device issue. Customers are to obtain the serial number of their system; verify the serial number obtained against the serial number listed in the box in the Appendix. A Maquet Service Representative will be contacting you to arrange a conveni
Quantity1641 [80 units (US) 1561 units (OUS)]
DistributionClass I Recall - Worldwide distribution -- US, including the states of AL, AZ, DC, FL, MI, MO, NJ, NY, OR, PA, TX, and WV; and, countries of Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Chile, China, Colombia, Croatia, Czech Republic, Ecuador, Finland, France, Germany, Hong Kong, Hungar
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133958FLOW-I ANESTHESIA SYSTEM C20, C30 & C40
Maquet Critical Care AB, Mar 20, 2014