Recall Z-1661-2015 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Siemens Medical Solutions Diagnostics

Direct Water feed Kit, Catalog#10373222

DateMay 23, 2015 (initiated Mar 5, 2015)
StatusTerminated
ReasonThe bottle assemblies were manufactured with a quality issue where in some cases the smaller diameter opening on both IMMULITE 2000 and IMMULITE 2000XPi Water Bottles and Liquid Waste Bottles is undersized and/or deformed, preventing the cap from closing or fastening securely to the bottle.
Root causeNonconforming Material/Component
ActionSiemens Healthcare Diagnostics sent an Urgent Device Correction Letter/Field Correction Effective Check Form dated 3/5/15 via FedEx to their affected customers.
Quantity15 bottles
DistributionWorldwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K970227IMMULITE 2000 AUTOMATED IMMUNOASSAY ANALYZER
Dpc Cirrus, Apr 8, 1997