Recall Z-1661-2015 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Siemens Medical Solutions Diagnostics
Direct Water feed Kit, Catalog#10373222
| Date | May 23, 2015 (initiated Mar 5, 2015) |
| Status | Terminated |
| Reason | The bottle assemblies were manufactured with a quality issue where in some cases the smaller diameter opening on both IMMULITE 2000 and IMMULITE 2000XPi Water Bottles and Liquid Waste Bottles is undersized and/or deformed, preventing the cap from closing or fastening securely to the bottle. |
| Root cause | Nonconforming Material/Component |
| Action | Siemens Healthcare Diagnostics sent an Urgent Device Correction Letter/Field Correction Effective Check Form dated 3/5/15 via FedEx to their affected customers. |
| Quantity | 15 bottles |
| Distribution | Worldwide Distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K970227 | IMMULITE 2000 AUTOMATED IMMUNOASSAY ANALYZER Dpc Cirrus, Apr 8, 1997 |